Emerson Process Management BINOS 100 Welding Consumables User Manual


 
PREPARATION
5 - 13
ETC00781(4) Series 100 e 02/2004
5.4.2 Electrical Connections
Be sure to observe the safety precautions and warnings !
Verify, that the cable glands together with installed cables are hermetic
to ensure protection class IP 65 (according to IEC 60529).
The permissible outside diameters of the cables are 7 to 12 mm !
a) Mains Supply
The analyzer is rated for an operating voltage of 230 V AC or 120 V AC, 47-63 Hz.
The built-in power supply is either of type SL5 or of type SL10 and provides a switch to
manually select the nominal voltage.
P Open the analyzers front door.
P Insert the mains line through the cable glands (Fig. 5-6 or 5-5) into the housing.
Connect L , N and PE to the terminals located nearby the powerline filter (Fig. 5-8).
Verify that the line voltage given on the identification plate
(front door inside) complies with that of your power supply line !
Verify that the position of the power supply's voltage selector switch complies
with the voltage of your power supply line (Fig. 27-7, 27-8 and 5-8) !
The BINOS
®
100 F (field housing) has no switch with disconnect function.
The customer has to provide a switch or circuit breaker into his installation. This
switch has to be installed near by analyzer, must be easily attainable for operator
and has to be charaterized as disconnector for analyzer.
Fig. 5-6: BINOS
®
100 F, cable glands for lines (side view from left)
ADDITIONAL HINTS ON BINOS
®
100 F (FIELD HOUSING)
Gas connections
Front door
Wall mounting
holder
cable gland
for mains line
cable glands
for data lines