Emerson Process Management BINOS 100 Welding Consumables User Manual


 
TECHNICAL DATA
27 - 6
ETC00781(4) Series 100 e 02/2004
27.4 Gas Related Specifications
For information about measuring components and ranges of the actual analyzer refer to the
nameplate label, which is located on the left or right side of the analyzer cover (all table top
and rack mount versions) or inside the front door (BINOS 100 F).
See chapter 27 for information about how to read a name plate label.
Available Measuring Components and Ranges
(some components / ranges may not be available with all analyzer versions)
Gas Component Minimum Ranges Maximum Ranges
Argon Ar 0 - 50 % 0 - 100 %
Carbon dioxide CO
2
0 - 100 ppm 0 - 100 %
Carbon monoxide CO 0 - 200 ppm 0 - 100 %
Ethylene C
2
H
4
0 - 400 ppm 0 - 100 %
Helium He 0 - 10 % 0 - 100 %
Hexane C
6
H
14
0 - 500 ppm 0 - 9,000 ppm
Hydrogen H
2
0 - 5 % ( 2 %
***)
) 0 - 100 %
Methane CH
4
0 - 2 % 0 - 100 %
n - Butane C
4
H
10
0 - 800 ppm 0 - 100 %
Oxygen O
2
0 - 1 %
***)
0 - 100 %
****)
Propylene C
3
H
6
0 - 4,000 ppm 0 - 100 %
Propane C
3
H
8
0 - 1,000 ppm 0 - 100 %
Toluene C
7
H
8
0 - 5,000 ppm 0 - 1,2 %
*)
Other components and configurations on request
**)
Dew point must not exceed ambient temperature
***)
Non-standard specifications