Emerson Process Management BINOS 100 Welding Consumables User Manual


 
1 - 1
TECHNICAL DESCRIPTION
ETC00781(4) Series 100 e 02/2004
1. Technical Description
The different analyzers are based on the same internal main components and differ only by
available options and enclosures.
BINOS
®
100 (M), OXYNOS
®
100, HYDROS
®
100 (1/4 19" enclosures)
BINOS
®
100 2M (1/2 19" enclosure )
All analyzer components are installed into enclosure, 3 height units tall. These housings are
classified protection class IP 20. The housings are available as rack-mountable or as table-top
versions. The table-top housings are fitted with an additional carrying handle and additional
rubber feets.
BINOS
®
100 F (field housing, wall mountable)
Additional we can deliver a field housing version . All components are installed into a protective
enclosure, classified IP 65 acc. IEC 60529 (approx. NEMA 4/4X). This enclosure is designed
for wall mounting. An magnetically operated impact tested front panel is available as an option.
Special versions are available for installation in hazardous areas, which meet the
requirements of either ATEX (Europe), CENELEC (outside Europe) or North America
(CSA-C/US; Z-purge).
The special conditions for operating ATEX analyzers are described in a supplemental manual.
1.1 Front Panel
The front panel includes the LED displays for both analysis channels and all of the analyzer
operating controls.
The BINOS
®
100 2M and BINOS
®
100 F front panels show status LEDs for the options
“Solenoid Valves” and “Gas Sampling Pump” and include a key “PUMP”
*)
to switch on and off
the gas sampling pump.
The front panels of BINOS
®
100 2M and BINOS
®
100 F may be equipped with an optional fine
dust filter with integrated needle valve or/and a flow meter (for general purpose applications
only).
These options are not available if BINOS
®
100 F is intended to be used in hazardous areas (EX
Zones) or if IP 65 is required.
FRONT VIEW