Emerson Process Management MMI-20014120 Welding Consumables User Manual


 
24 Micro Motion 3098 Gas Specific Gravity Meter
Electrical Connections
Figure 3-9 7950 signal converter and gas specific gravity 3-wire system with shunt-diode safety barrier
(hazardous area)
Figure 3-10 7950 signal converter and gas specific gravity 3-wire system with galvanic isolator (hazardous
area)
Note: The barrier trip level switch should be set to 3 volts.
Note: When the ATEX/IECEx-approved specific gravity meter is installed in a hazardous area, the
safety instruction booklet shipped with the unit is the authoritative document.
3.5.3 7951 2-wire configuration
Figure 3-11 7951 flow computer/7951 signal converter gas specific gravity 2-wire system (safe area)
3098 meter
1
+
SIG A
2
-
3
+
SIG B
4
-
Hazardous Area Safe Area
MTL 787 (+ve)
3
4
1
2
7950 Flow Computer/Signal Converter
Density power +PL10/1
Ch.3
PL10/2
PL10/3
PL10/4
PL10/5
Ch.4
PL10/6
PL10/7
PL10/8
Density input +
Density input -
Density power -
Meter
1
+
SIG A
2
-
3
+
SIG B
4
-
MTL 5532
4
1
14
13
Hazardous Area Safe Area
Density power +PL10/1
Ch.3
PL10/4
PL10/3
PL10/2
PL10/5
Ch.4
PL10/8
PL10/7
PL10/6 Density input +
Density input -
Density power -
7950 Flow Computer/Signal Converter
5 12
11
2kR
3098 meter
1
+
SIG A
2
-
3
+
SIG B
4
-
330R
7951 Signal Converter/Flow Computer
24V pwr +
PL5/9 (SK6/22)
Ch.3
PL5/5 (SK6/18)
PL5/6 (SK6/19)
PL5/10 (SK6/24)
PL5/9 (SK6/22)
Ch.4
PL5/7 (SK6/20)
PL5/8 (SK6/21)
PL5/10 (SK6/24)
Den ip +
Den ip -
24V pwr -
(0V dc)
(Den -)
(+24V dc)
(Den +)