Fuji Labs ZSP Welding Consumables User Manual


 
35
10.6 Maintenance of zirconia O
2
sensor
Sensor temperature
800
OXYGEN ANALYZER
POWER
ON
OFF
1234567
Power SW Fuse 3A Terminal
Front view Side view
Sample inlet
(Rc 1/4)
Outlet (Rc 1/4)
AG 0V
+
ALM
(1) Appearance of zirconia O
2
sensor
(2) Routine maintenance
No parts require periodic maintenance. In case sensor temperature is 800±5°C or the SO
2
concentration in
sample gas is high, you should check occasionally that the outlet pipe is not clogged by the precipitation of
crystal. Because the sample outlet is at a high temperature, attention should be paid not to suffer a burn.