6 - 10
No. 99MBC032A
6.1.6 List of response commands if an error occurs
Response command Description
A workpiece is not present in the specified segment.
• A workpiece is not set properly.
• Shutter is closed.
Parity error occurred.
• Check the setup contents in the basic setup.
• Isolate the cables from noise sources.
All setup data are initialized (cleared) when the power is ON.
A numeric value greatly different from the reference gage dimension is set.
• Limit values for go/no-go judgment and abnormal data exclusion have been set in reverse
order or equal.
• Input value is too large.
An unavailable command is received.
• Command format is incorrect.
• Baud rate and/or data bits are not consistent.
Message from the external device
• Measurement is interrupted by signal input from key operation.
ER0
ER1
ER2
ER9
ER5
ER6
ER7