No. 99MBC032A
4. SETTING UP THE MEASURING CONDITIONS
4.1.2.3 Selecting and setting the function in the B2 mode
a. Setting the workpiece type (Guidance: )
Set whether the workpiece is an opaque object or transparent object.
: Workpiece is an opaque object.
: Workpiece is a transparent object.
(Default setting: )
PROG
Step 1: Each time the key is pressed the displayed
setup option toggles between and
. While the desired setup option is
flashing, press the
ENT
key. The operation
automatically proceeds to setting whether to
perform simultaneous measurement.
TIP If is selected for the workpiece type, the guidance for the selection of the
segment specification method is not displayed. It is omitted (the segment specifica-
tion process is entered directly).
b. Setting the simultaneous measurement (Guidance: )
Set whether to perform simultaneous measurement.
: Does not perform simultaneous measurement. (performs single measurement)
: Performs simultaneous measurement.
(Default setting: )
PROG
Step 1: Each time the key is pressed the displayed
setup option toggles between and
While the desired setup option is
flashing, press the
ENT
key.
TIP If (simultaneous measurement) is selected
If simultaneous measurement is selected, the setup guidance for the following will
not be displayed: Selecting the averaging method, segment specification, and setting
the group judgment.
c. Selecting the method of specifying segments (Guidance: )
Select the method of specifying the measurement position from segment specification and
edge specification.
: Uses segment specification.
: Uses edge specification.
(Default setting: )
PROG
Step 1: Each time the key is pressed the displayed
setup option toggles between and
. While the desired setup option is
flashing, press the
ENT
key. The operation
automatically enters B3: Setting the abnormal
value elimination function.
NOTE If any of the following setting is performed, the system automatically proceeds to the
stage of segment setup without displaying the guidance for the
method of specifying segments SEG:
• a. is selected in Setting the workpiece type.