4 - 34
No. 99MBC032A
4.5.3.6 F5: Setting the data output conditions
Set the data output conditions (0 to 9) and periodical output timer (0 to 999 sec.).
The unit used with the periodical output timer is seconds. Setting it to “0” means that output
takes place for each measurement.
In this example assume that the data output condition is 3 and that the periodical output timer
is 10 seconds.
PROG
Step 1: The previously set data output conditions are
displayed.
The setup data for the data output conditions is
shown in the table below.
C.RUN
RUN
C.RUN
RUN
C.RUN
RUN
Data output condition Printer
RS-232C
DCU
Remark
The periodical output timer can be set
0
1
2
7
——
—
—
—
—
8
9
6
The periodical output timer can be set5
4
The periodical output timer can be set3
: Outputted for each measurement if or key, etc., is pressed.
: Press the or key to trigger the measurement. The measurement result will be outputted if it
falls on GO.
: Press the or key to trigger the measurement. The result will be outputted if it falls on ±NG.
: No output will be made.
—
—
—
3
PROG
Enter “3” as the data output condition.
PROG
Step 2: Press the
ENT
key.
If the data output condition is 1, 3, or 5, the
operation proceeds to the periodical output timer
setting, otherwise it proceeds to the sample
measurement setting.
Step 3: Set the periodical output timer to 10 seconds.
10
PROG
PROG
Step 4: Press the
ENT
key to save the setup data in
memory.
The operation enters the sample measurement
setting.