6 - 18
No. 99MBC032A
(16) R
(a) Format: R
(b) Description: If the number of samples is set between 1 and 999, this command executes
single-run measurement and transmits the measurement result in conformity
with DATA FORMAT as the response command.
(c) Example: Reception command R
Response command P0, 12.3456
(17) CR, CL
(a) Format: CR
CL
(b) Description: CR: If the number of samples is set between 1 and 999, this command executes
continuous-run measurement. However, it does not respond to the “CR”
command.
• It transmits the measured results in conformity with DATA FORMAT for
the response command.
CL: Terminates continuous-run measurement.
(c) Example: Reception command Response command
CR None
• Outputs as the response, the measurement results according to the data
output conditions in conformity with DATA FORMAT.
CL OK
(18) R, STOP
(a) Format: R
STOP
(b) Description: R: If the number of samples is set to 0, this command executes zero-run
measurement. However, it does not respond to “R” command.
STOP:Terminates the zero-run measurement, and transmits the measurement
results in conformity with DATA FORMAT as the response.
(c) Example: Reception command R
STOP
Response command P0, 12.3456
(19) D
(a) Format: D
(b) Description: Transmits as the response the last display of data in the ready state or latched
data not in conformity with DATA FORMAT.
This command is used to transmit the previous data, while the R command is
used to execute measurement then the results are transmitted.
(c) Example: Reception command D
Response command (DATA FORMAT)