4 - 24
No. 99MBC032A
2) Edge specification
PROG
Step 1: If the edge specification mode is entered, the
previously established manual measurement/
automatic measurement item will be displayed.
PROG
Each time the key is pressed the setup
option changes in the following order:
Manual measurement:
→ Automatic
measurement for pitch:
→ Automatic
measurement for diameter: → Automatic
measurement for gap: . If the desired setup
option is displayed, press the
ENT
key.
Operation automatically enters the process of setting the start edge.
PROG
2
Step 2: Set the start edge (between number 1 and
254)
In this example, set the start edge to
number 2.
If the
ENT
key is pressed, the operation
automatically enters the process for setting
the finish edge.
PROG
5
6
Step 3: Set the end edge (between number 2
and 255)
In this example, set the end edge to
number 65.
If the
ENT
key is pressed, the operation auto-
matically proceeds to F1: Setting the measurement
interval.
TIP 1. If a calibration is performed using the reference gage (placed in SEG2) with the
edge specification active, select Manual measurement: or Automatic
measurement for diameter: , then set the start edge to 2 and end edge to 3.
2. If the checks on the start and end edges, performed at the end of the setup
operation, result in start edge > end edge, exchange the start and end edge data.
If the check result shows that both edge numbers are identical, an error (Err-5)
results. If this occurs, cancel the setup data and begin the setting with the start
edge.