4 - 16
No. 99MBC032A
(4) Setting the HIGH CAL gauge.
HIGH CAL gages vary in shape depending on the LSM model to be calibrated. Set the
calibration gage so that the calibration guide line ( | ) on the side face of the calibration
gage comes vertical, and so that the center of the calibrated section is measured.
In diagram (a), the calibrated position is at the center of the ( | | ) mark, and the center of
the width (indicated by the arrow mark) in diagram (b).
(a) (b)
Step 1: Cancel the previously set calibration values.
It is not necessary if this setup operation is made with the previously used calibration
gage. However, if the new gage diameter is much different from that of the previous
one, an error (Err-2) may result. If this is the case, cancel the LOW CAL calibration
value, then begin with the setting of HIGH CAL value (it does not matter if both the
LOW CAL and HIGH CAL values are canceled).
PROG
1) Cancel the previous LOW CAL data. Press the
L.CAL
key in the ready state to initiate the
LOW CAL setup mode.
PROG
C
PROG
ENT
2) Press the
C
and
ENT
keys to
cancel the LOW CAL data. This
automatically restores the ready state.
PROG
Step 2: Mount the HIGH CAL gage on the stand.
Press the
H.CAL
key in the ready state. The
previously set HIGH CAL value is displayed, and
the HIGH CAL setup mode is entered.
Step 3: Enter the approved dimension of the HIGH CAL
gage.
Example.)
PROG
2
00
1
24
.