3 - 34
No. 99MBC032A
PROG
8) Press the
ENT
key to save the setup value in
memory.
IMPORTANT Rounding setup value
Setup value will be rounded off automatically if its least significant digit does not
agree with the resolution of the display.
Example: In case the resolution is 0.05 μm
345,64 > 12.345,60 (least significant digit 4 is rounded off to 0)
345,67 > 12.345,65 (least significant digit 7 is rounded off to 5)
TIP About the input of setup data
PROG
1. How to enter a sign
If “Perform GO/NG judgment by (target value +
tolerance)” has been specified in the basic setup
and the lower tolerance limit is “-0.015”, input as
follows. In this case a “0” does not need to be
placed in the integer section.
PROG
00 1 5
.
)
PROG
+/
-
2.
READ
key: About the read operation
Generally, in the calibration or offset value setup operation a reference gage is
used, resulting in a measured data that is very close to the setup value. If this is
the case, first read a measurement as the setup data, then correct the minor
difference.
3. To enter a numeric value such as a gage diameter, it is more convenient to use
the numeric keys. To correct a specific digit, it is more convenient to use the
arrow keys.
4. To select a setup item such as the resolution in the basic setup, it is better to use
the or key. Use of a numeric key causes an operation error.