Chapter 2 9
Setting up the HP/Agilent 8922 GSM HSCSD Test System
Making the rear panel connections
Making the rear panel connections
Before connecting the two HP/Agilent 8922 P systems to form the HP/
Agilent 8922 GSM HSCSD Test System, ensure that both HP/Agilent
8922 P systems are connected and calibrated as described in the HP/
Agilent 8922 Multi-Band Test System Supplementary User’s Guide. See
the HP/Agilent 8922 GSM Test Set User’s Guide and the HP/Agilent
83220 User’s Guide for details on the rear panel connections.
To make the rear panel connections of the HP/Agilent 8922 GSM HSCSD
Test System:
Step 1. Connect the system bus cable from the master unit to the slave unit (the
end with the BNC connection must be connected to the master unit).
If you are not connecting the BNC connector on the system bus cable to
for example, a spectrum analyzer, then terminate the connection using
the BNC terminator provided.
Step 2. Connect the GPIB cable from the GPIB port on the master unit to the
GPIB port on the slave unit.
NOTE If you are operating the HP/Agilent 8922 GSM HSCSD Test System
manually do not connect another GPIB controller to the GPIB bus. The
master unit operates as the controller and another device attached to the
GPIB bus results in conflicts on the GPIB bus.
Step 3. Connect the 10 MHz reference cable from the 10MHz OUT port on the
master unit to the Ref IN port on the HP/Agilent 83320 of the slave unit.
The HSCSD connections are shown below.