Agilent Technologies 8922 P Welder User Manual


 
16 Chapter3
Manual operation
Configuring the system
Configuring the system
To configure the HP/Agilent 8922 GSM HSCSD Test System for manual
operation, proceed as follows.
NOTE You must configure the slave unit first as a frame reset is sent from the
master unit to the slave unit as soon as the master unit is activated. The
slave unit must be ready to receive this frame reset.
NOTE If you are operating the HP/Agilent 8922 GSM HSCSD Test System
manually do not connect another GPIB controller such as a PC to the
GPIB bus. The master unit operates as the controller and any other
device attached to the GPIB bus can result in conflicts on the GPIB bus.
Step 1. On the slave unit, navigate to the CONFIG screen and set the Instrument
Type to HSCSD Slave.
Figure 3-1 Config screen
C.01.06