Agilent Technologies 8922 P Welder User Manual


 
50 Chapter5
Remote operation
HSCSD Subsystem GPIB Commands
HSCSd:MASTer:DOWNlink:AMPLitude
Description This sets the downlink amplitude of the master TCH. In all cases, this
sets the TCH level. To change the amplitude of the GSM BCH in a dual
band system, use the existing command DUALband:ATTenuation:GBCh.
See the HP/Agilent 8922 Multi-Band Test System Supplementary User’s
Guide for more details.
This command must also be sent to the slave unit.
Syntax HSCSd:MASTer:DOWNlink:AMPLitude?
HSCSd:MASTer:DOWNlink:AMPLitude <real>
Command 0ptions
Type floating point
Units: dBm only
Limits (GSM900): -127.0 thru +1.0
Limits (E-GSM): -127.0 thru +1.0
Limits (DCS1800): -127.0 thru -12.0
Limits (PCS1900): -127.0 thru -12.0
Default: -85.0