Chapter 4 25
Screens
Config screen
set the slave GPIB address to match the address of the slave unit
make sure that the addresses of the slave and master units are
different
make sure that no other GPIB controller is active and attached to the
system
NOTE If you are operating the HP/Agilent 8922 GSM HSCSD Test System
manually do not connect another GPIB controller such as a PC to the
GPIB bus. The master unit operates as the controller and any other
device attached to the GPIB bus can result in conflicts on the GPIB bus.
Selecting HSCSD Slave sets the GPIB address of the instrument to 15 to
match the default slave GPIB address of the master unit.
If the HP/Agilent 8922 GSM HSCSD Test System is not fitted with a new
HOP Controller ROM, selecting HSCSD Master or HSCSD Slave results
in the error Hardware portion of upgrade not present. The test set
will allow HSCSD operation, however the downlink signal could be
scrambled when the TCH ARFCN is on or near the master or slave unit’s
BCH ARFCN. For details on upgrading the HP/Agilent 8922’s HOP
Controller ROM, see “Upgrading the HOP Controller ROM and
Firmware” on page 75.
NOTE The Instrument type field is reset to the default in the event of a reset
or power cycle. Setting this field also resets all the default values on the
HSCSD Control screen.