36 Chapter4
Screens
HSCSD screen
On both units, if the instrument is operating in standard mode, setting
this field changes the slave control field (see page 37) to MS Tx Lev, then
calculates and sets the slave expected input level (see page 36). If the
instrument is operating in compatibility mode, setting this field only
changes the slave expected input level when the slave control type is MS
Tx Lev.
NOTE The default is reset when you select the instrument type (see page 24).
13. Slave input level
The Slave input level field sets the expected power of the slave uplink
TCH. You can set the level indirectly by changing the Slave Tx level
(see page 35) or change it directly. This value is used to set the RF
analyzer level when the measurement timeslot (see page 32) is set to
Slave.
If the instrument is operating in standard mode, setting this field
changes the slave control type (see page 37) to Manual.
If the instrument is operating in compatibility mode, and the slave
Type float
Units dBm only
Limits (GSM900) -27 thru +41.0
Limits (E-GSM) -27 thru +41.0
Limits (DCS1800) -47 thru +33.0
Limits (PCS1900) -47 thru +33.0
Default (compatibility mode) 33.0
Default (GSM900) 13.0
Default (E-GSM) 13.0
Default (DCS1800) 10.0
Default (PCS1900) 10.0