Chapter 5 43
Remote operation
Configure Subsystem GPIB Commands
Configure Subsystem GPIB Commands
Figure 5-1 Configure Instrument Type comand syntax
CONFigure:INSTrument[:TYPE]
Description Sets what role the HP/Agilent 8922 plays in the multi-slot test system.
This selection differs from manually selecting the instrument type in two
ways: the GPIB status does not change to Control, and the unit’s GPIB
address does not change.
Note that you need to manually select two different GPIB addresses for
the master and slave units.
If the instrument has not been fitted with a new hop controller ROM,
selecting HSCSD Master or HSCSD Slave results in the error Hardware
portion of upgrade not present. In this case the instrument allows
HSCSD operation, but the downlink signal could be scrambled when the
TCH ARFCN is placed on or near to the master or slave's BCH ARFCN.
Syntax CONFigure:INSTrument[:TYPE]?
CONFigure:INSTrument[:TYPE] <string>
Command options ‘SPEECH’ | ‘HSCSD MASTER’ | ‘HSCSD SLAVE’
CONFigure
INSTrument[:TYPE]
Sp
string
INSTrument[:TYPE]?