SERIAL INTERFACE (OPTION)
12 - 13
ETC00781(4) Series 100 e 02/2004
12.5.2 Response Telegrams
Response telegrams follow with the same syntax as the appropiate (SP-) commands (see 12.5).
12.5.2.1Status Messages (Instruction 030)
The response telegram for instruction
“$ID;030;LPB<CR> SP Status messages”
shows as follows:
$ID;030;a;b;c;LPB<CR>
This means:
a: OK-Status 0 = Relay without power 1 = Relay active
b: Value of variable “calibration” 0 = Relay without power >0 = Relay active
b Meaning
0 No Calibration
1 Zeroing channel 1
2 Zeroing channel 2
3 Zeroing channel 1 + 2
4 Spanning channel 1
5 Spanning channel 2
6 Spanning channel 1 + 2
7 Spanning channel 1 first, then channel 2
8 reserved
9 reserved
10 Waiting for flushing time and t
90
response time
c: Gas sampling pump
*)
0 = Relay without power 1 = Relay active
(pump OFF) (pump ON)
*)
BINOS
®
100 2M/F from program version 5.0 is required
RESPONSE TELEGRAMS