Emerson Process Management HYDROS 100 Welding Consumables User Manual


 
Error Code Possible Reasons Check / Correct
16 - 1
LIST OF FAILURES
ETC00781(4) Series 100 e 02/2004
16. List of Failures
Some of the failures which may arise during measurement will be reported on the displays in forms
of error codes.
When such a failure arises, the display's will show the concentration value
alternating with (E = ERROR).
Note !
If there is an "error message", a digital status signal "Failure" can be given optional (see Item 10.3)!
Be sure to observe the safety measures for all workings at the analyzer!
1. Press any key.
Check parameter dOFF
(see chapter 8.7).
2. Check connection mains line/PS
Check external power supply
Check electrical supply
(see Fig. 1-5 and 1-6, Item 3
Fig. 1-7 to 1-9, Item 10).
Check internal fuses F1 and F2
3. Check connection
BKB - BKS (X1) (see chapter 18.1).
1. Displays are “switched OFF”
2. Voltage supply absent.
BINOS
®
100 F
3. Connection front panel /BKS
absent.
No Display