Emerson Process Management HYDROS 100 Welding Consumables User Manual


 
PREPARATION
5 - 8
ETC00781(4) Series 100 e 02/2004
5.3.2 Internal Solenoid Valves (BINOS
®
100 2M/F Option)
For operation with optional internal solenoid valves, the following indications have to be
considered:
All necessary gases have to be connected at the corresponding solenoid valve at an overpressure
of 50 - max. 500 hPa.
For 2 channel analyzer with parallel gas paths we can deliver a special solution with 2 valve blocks.
Be sure to observe the safety regulations for the respective gases
(sample gas and test gases / span gases) and the gas bottles !
The test gas containers have to be set up according to the valid regulations.
If a solenoid valve is open there is illuminated a green LED (Fig. 1-2, Item 3) at the front panel.
Fig. 5-4a: BINOS
®
100 2M (version A), gas connections with solenoid valve option
GAS CONNECTIONS
Solenoid valve:
common gas outlet
(to standard gas inlet K1)
Solenoid valve:
span gas 1
Solenoid valve:
span gas 2
Solenoid valve:
sample gas
Solenoid valve:
zero gas