Emerson Process Management HYDROS 100 Welding Consumables User Manual


 
MEASURING POINTS OF BKS AND OXS
17 - 3
ETC00781(1) Series 100 e 10/2001
Rosemount Analytical
17.1.5 Temperature Sensor
Measuring point: X 8
Measuring device: DVM
Signal: approx. 0 ± 500 mV DC (at ambient temperature)
Failure. Signal > + 3,5 V DC
Possible reasons (IR measurement or paramagnetic oxygen measurement):
a) Temperature sensor not connected (see chapter 18.1).
b) Temperature sensor faulty (exchange sensor).
c) Broken cable of temperature sensor (exchange sensor).
d) BKS faulty (exchange BKS).
Possible reasons (electrochemical oxygen measurement):
a) Temperature sensor not connected (see chapter 18.2).
b) PCB OXS faulty (exchange OXS).
d) PCB BKS faulty (exchange BKS).
MEASURING POINTS OF BKS