5.1 Installation Site
The analyzers must not operate in explosive atmosphere without supplementary
protective measures !
Free flow of air into and out of the analyzers (ventilation slits) must not be
hindered by nearby objects or walls !
Operate analyzer as table-top version or as rack-mountable version (built-in) only
(except of BINOS
®
100 F: designed for wall mounting only) !
The installation site for the analyzers has to be dry and remain above freezing
point at all times. The analyzers must be exposed neither to direct sunlight nor
to strong sources of heat.
Be sure to observe the permissible ambient temperatures (c.f. chapter 27:
Technical Data). For outdoor installation, we recommend to install the analyzers
in a protective cabinet. At least, the analyzers has to be protected against rain
(e.g., shelter).
The BINOS 100 series analyzer has to be installed as near as possible to the sample point,
in order to avoid low response time caused by long sample gas lines.
In order to decrease the response time, a sample gas pump with a matching high pumping rate
may be used, while the analyzers has to be operated in the bypass mode or by an overflow
valve to prevent too high flow and too high pressure (Fig. 5-2).
Fig. 5-2: 100 series, Bypass installation
INSTALLATION SITE
Overpressure valve
Flow meter
Exhaust
Exhaust
analyzer
Gas sampling pump
Filter