Fluke 7350 Air Compressor User Manual


 
Local Operation
Menu 4
4-15
Menu | Setup — Remote
The Menu | Setup | Remote screen is where the remote communication interface
is set-up.
gld16.bmp
Figure 4-14. Menu | Setup | Remote - Menu
GPIB Address
Sets the IEEE-488 interface address.
Protocol
The Protocol defines which protocol should be used by the remote interface. The options
are Standard Communication for Programmable Instruments (SCPI) which is the standard
interface, 6000 to emulate the Series 6000 pressure controller, and 510 to emulate the
Model 510 pressure controller. When using the SCPI interface, the RUSKA 7350 will
emulate the 7010 and 7215 series controllers.
Serial Interface Set-Up
The operator can set-up the Baud Rate, Data Bits, Parity, and Stop Bits for the Serial
Interface.
Menu | Setup — System
The System Menu identifies the Software release version that is being operated in the
Calibrator. It also stores and allows the operator to edit the Date and Time.
gld17.bmp
Figure 4-15. Menu | Setup | System - Menu