Fluke 7350 Air Compressor User Manual


 
Remote Operation
Device Messages 5
5-5
:MODE? calibration edit enabled?
:MODE ON¦OFF¦1¦0 enable calibration edit
(Cal. button required)
DISP
:ENABle ON¦OFF¦1¦0 turns front panel display on/off
:TEXT <string> displays message on front panel
:BGRaph <number> sets bar graph maximum
OUTPut
:STATe ON¦OFF¦1¦0 off=MEASure, on=CONTrol
:STATe? returns 0=Measure or 1=Control
:MODE MEASure¦CONTrol¦VENT sets mode
:MODE? returns mode string
PROGram
:CATalog? returns list of defined programs
[SELected]
:DEFine <program block> Define program
press1, toler1, dwell1, max1, press2,
toler2, ..
:DEFine? read program definition
:DELete
[:SELected] deletes current program
:ALL deletes all programs
:NAME <program name> select current program
:STATe RUN|PAUSe|STOP|CONTinue set program state
:STATe? ` read program state
:CONFigure
:RECall restore saved configuration
:SAVE save current configuration
SENSE
[:PRESSure]
[:RESolution] <number> set pressure display resolution
:AUTO <boolean>|ONCE return to default resolution
:MODE? returns ABSOLUTE or TARE
:RANGE
[:UPPer] returns RUSKA 7350 full scale value in
units
:LOWer returns 0
:REFerence
[:HEIGht] <number> set head height
:SGRavity <numbers> set specific gravity
[SOURCE]
[:PRESsure]
[:LEVel]
[:IMMediate]
[:AMPLitude] <number> sets pressure setpoint
[:AMPLitude]? read pressure setpoint
:MODE FIXed|LIST set source parameter set
:TOLerance <number> specifies output tolerance
:SLEW <number> set slew rate
:CONTrol <number> set control band