Fluke 7350 Air Compressor User Manual


 
RUSKA 7350
Users Manual
5-4
ANSI/IEEE 488.2-1987 Command Summary
*CLS Clear Status
*ESE? Event Status Enable Query
*ESE <number> Event Status Enable
*ESR? Event Status Register
*IDN? Identification
*OPC? Operation Complete Query (Returns 1)
*OPC Operation Complete
*RST Reset
*SRE? Service Request Enable Query
*SRE <number> Service Request Enable
*STB? Status Byte Query
*TST? Self-Test Query
*WAI Wait (No operation)
SCPI Command Summary
The current value associated with a SCPI command may be read by appending a question
mark to the command. For example CALC:LIM:UPP? will return the current upper
pressure limit.
MEASure
[:PRESsure]? returns current pressure reading
:POSITION? returns pump position (%)
CALCulate
:LIMit
:LOWer <number> get/set low pressure limit
:SLEW <number> get/set slew rate limit
:UPPer <number> get/set high pressure limit
:VENT <number> get/set auto-vent limit
:TARE
:VALue <number> get/set tare value
:STATe ON/OFF set tare using current pressure
CALibration
[:PRESsure]
:Data
:POINts ? number of calibration constants
:VALue<n>? returns cal constants label, value
:VALue<n> <number> set calibration constant
:CALibration
:POINts? number of calibration points
:VALue? Nominal calibration point
:VALue<n> <number> perform calibration point
:ZERO performs zero calibration
:VALUE <number> sets zero value
:INITiate enter zero calibration mode
:INITiate? returns status for cal, press, temp, ref
:RUN start zero calibration
:STOP abort zero calibration
:POSition
:DATA <number>, <number> sets C0, C1 position
:DATA? reads C0, C1
:VALue <number> sets first calibration point
:VALue 2 <number> sets second calibration point