Fluke 7350 Air Compressor User Manual


 
5-1
Chapter 5
Remote Operation
Remote Operation
The RUSKA 7350 can be operated remotely by a computer. Three interfaces are
supported; IEEE-488, RS-232, and RS-485. All three interfaces support Standard
Commands for Programmable Instruments (SCPI).
The IEEE-488 interface additionally supports emulation of a Single Channel Interface
Panel (Models 6005-701 and 6005-761). The IEEE-488 interface conforms to the
following standards.
ANSI/IEEE Std 488.1-1987 IEEE Standard Digital Interface for Programmable
Instrumentation
ANSI/IEEE Std 488.2-1987 IEEE Standard Codes, Formats, Protocols, and Common
Commands
SCPI 1991.0 Standard Commands for Programmable Instruments
Capabilities
IEEE-488
The following identification codes define the interface capabilities of the RUSKA 7350.
Their meaning is described in the IEEE-488 standard.
SH1 Source Handshake, Complete Capability
AH1 Acceptor Handshake, Complete Capability
T5 Talker
L3 Listener
SR1 Service Request, Complete Capability
RL1 Remote-Local, Complete Capability
PP0 Parallel Poll, No Capability
DC1 Device Clear, Complete Capability
DT0 Device Trigger, No Capability
C0 Controller, No Capability
The optional IEEE-488 interface is installed next to the processor board. The interface is
identified by the IEEE-488 standard connector on the back panel of the unit.