Fluke 7350 Air Compressor User Manual


 
RUSKA 7350
Users Manual
4-24
4. Press Run [F2]. The configuration of the Calibrator stored with the program is
restored, the pressure set-point is set to the pressure value in the first step, and the
Calibrator is placed in Control mode. “Run” will now be highlighted and the program
will proceed through its steps.
5. To pause the program, press Pause [F3]. “Pause” will now be highlighted and the
Calibrator will continue controlling to the current set-point. The Calibrator will
maintain control of the current set-point until further instructions from the operator.
Press Continue [F4] to resume the program.
6. To stop the program, press Stop [F5]. The program will stop running but the
Calibrator will continue controlling to the current set-point.
Menu | Test
The Menu | Test section of the system is used to perform a number of the diagnostic and
tuning functions that are available with the RUSKA 7350.
gld25.bmp
Figure 4-23. Menu | Test - Menu
Menu | Test — Sweep Test
The sweep function can be used to automatically exercise the elastic sensing element of
the device under test prior to performing a calibration. For the Calibrator to perform the
sweep test, input the high and low set-point pressures, the pressure control tolerance, the
dwell time at set-point, and the number of cycles to perform.
1. Ensure that the units, limits, and control parameters are set to their desired values.
2. The sweep test is entered from the Sweep Test Menu. From the Main Menu, (press
PREVIOUS until the Main Menu appears), press Menu [F6], then Test [F5], and then
Sweep [F2].
3. Enter the high and low pressure points, the control tolerance, dwell time, and the
number of cycles.