Fluke 7350 Air Compressor User Manual


 
RUSKA 7350
Users Manual
4-22
2. Use the rotary knob to highlight the name of the program.
3. Press Edit [F4]. The program editing screen will appear as a table showing all of the
steps.
gld22.bmp
Figure 4-20. Menu | Program | Edit - Menu
4. The rotary knob can be used to move through the program and highlight the value
you wish to edit. Use the keyboard to enter a new value. This value will show up in
the scratch pad located at the bottom of the window. Press Enter to accept the
edited value.
5. To add a step to the program, first move to the step after the new step. For example,
to insert a step between steps 3 and 4, move to step 4. Press Insert [F4].
6. To delete a step in the program, first move to the step to be deleted, then press
Delete [F5].
7. When all changes have been made, press PREVIOUS to return to the Named
Programs screen. The Calibrator automatically saves the changes that were made in
the program.
Changing the Configuration Stored with a Program
1. Programs are configured from the Program Menu. From the Main Menu (press
PREVIOUS until the Main Menu appears), press Menu [F6] | Program [F4].
2. Use the rotary knob to highlight the name of the program.
3. Press Config [F3]. The configuration screen will appear.
4. Press Recall [F2]. The Calibrator will be set to the configuration stored with the
program.