Fluke 7350 Air Compressor User Manual


 
RUSKA 7350
Users Manual
4-20
Automatically Generating a Program
For the Calibrator to automatically generate a program, the user must input the first
set-point pressure, the last set-point pressure, and the number of steps in between, as well
as the dwell time, max time and tolerance common to all set-points.
1. Ensure that the units, limits, and control parameters are set to their desired values.
2. The program is entered from the Program Menu. From the Main Menu (press
PREVIOUS until the Main Menu appears), press Menu [F6] | Program [F4].
3. Use the rotary knob to highlight “new”. If these steps are used on an existing
program, all program steps will be deleted and replaced with the automatically
generated program.
4. Press Edit [F4]. Since “new” was highlighted, the Calibrator will create a new
program and give a default name of NAMEnn where nn is a two digit number. The
Program editing screen will appear, displaying the first step.
5. Press Auto [F3].
gld20.bmp
Figure 4-18. Menu | Program | Edit | Auto - Menu
6. Using the numeric keypad, enter the values for Start, Stop, Tolerance, Dwell time,
Max time, number of pressure Points up, and number of pressure Points down,
pressing ENTER after each value. The rotary knob may be used to skip fields.
7. Press Program [F1]. The program will be generated and the display will show the
first step.
8. Press the Name [F6] key to edit the name of the program. See next section, Chapter 4,
Changing the Name of a Program, for creating a user-defined program name.