Fluke 7350 Air Compressor User Manual


 
RUSKA 7350
Users Manual
5-6
:LIST
:PRESsure <number> [,<number>] set list of pressure values
:POINts? returns number of points defined
:DWELl <number> [,<number>] specifies dwell times
:POINts? returns number of dwell times
:TOLerance <number> [,<number>] specifies tolerances
:POINts? returns number of tolerances
:DIRection UP|DOWN direction to go through list
:COUNt <number> number of times to go through list
STATus
:OPERation
[:EVENT]? read/clear operation event register
:CONDition? read operation condition register
:ENABle <number> set operation enable mask
:QUEStionable
[:EVENT]? read/clear questionable event register
:CONDition? read questionable condition register
:ENABle <number> set questionable enable mask
:PRESet reset condition flags
SYSTem
:DATE <year>,<month>,<day> set system date
:ERRor? returns <error#,”descr;info”>
or 0,”No Error”
:KLOCk ON|OFF|1|0 lock keyboard
:TIME <hour>,<minute>,<second> set system time
:VERSion? returns 1991.0
:LANGuage “6000”|”SCPI” set interface protocol to 6000 or SCPI
:PRESet reset system
TEST
:ELECtronic? perform electronic self-test
UNIT
:DEFine<n> <name>,<number> define a unit
:LENGth MM|IN set length units for head height
[:PRESsure] <unit name> set pressure units
Example SCPI Commands
To request the current pressure reading, all of the following commands are equivalent:
:MEASURE:PRESSURE?
:measure:pressure?
:MeAsUrE:pReSsUrE?
:meas:pres?
:measure?
:meas?
MEAS?
To set the control pressure setpoint to 50, all of the following commands are equivalent:
SOURCE:PRESSURE:LEVEL:IMMEDIATE:AMPLITUDE 50
SOUR:PRES:LEV:IMM:AMPL 50.0
PRESSURE +50
PRES 50