Value Description and Corrective Action
-286 Program Runtime Error. An error occurred while running the program. Usually the
setpoint is out of range.
-313 Calibration Data Lost. The calibration data has been lost and the unit must be
recalibrated.
-315 Configuration Data Lost. The configuration data has been lost. Check all parameters
to be sure that they are correct.
-330 Self Test failed. See display for more information.
-350 Queue Overflow. The error queue was full and messages were lost.
-400 Query Error. A read request was received when there was nothing to read.
500 Controller Malfunction. Internal control failure.
501 High Limit Exceeded. The pressure was greater than the high limit.
502 Low Limit Exceeded. The pressure was less than the low limit.
503 Slew Limit Exceeded. The pressure changed faster than the slew limit allowed.
521 Pressure Overrange. The pressure reading is outside the range of the RUSKA 7350.
538 Auto vent. The Vent Limit was exceeded and the instrument entered Vent Mode.
543 Supply Pressure. The instrument is attempting to increase pressure and the
pressure is not responding. Check main pressure supply, shop air supply, and for
leaks.
600 Factory Data Lost. Internal factory constants have been lost. Contact Fluke for more
information.
602 Model Changed. Firmware has been downloaded that is not correct. Download the
correct firmware.
706 Pump position failure. The position sensor is not detecting movement in the pump
after the pump has been commanded to move.
707 Range Switch. The current range can only be changed when the pressure reading is
below 100 psi (700 kPa).