Agilent Technologies E4356A Welding System User Manual


 
Remote Programming 51
SCPI Data Formats
All data programmed to or returned from the power supply is ASCII. The data may be numerical or character string.
Numerical Data
Table 6-1 and Table 6-2 summarize the numerical formats.
Table 6-1. Numerical Data Formats
Symbol Data Form
Talking Formats
<NR1> Digits with an implied decimal point assumed at the right of the least-significant
digit. Examples: 273 0273
<NR2>
Digits with an explicit decimal point. Example: 273. .0273
<NR3>
Digits with an explicit decimal point and an exponent. Example: 2.73E+2 273.0E-2
Listening Formats
<NRf>
Extended format that includes <NR1>, <NR2> and <NR3>. Examples: 273 273.
2.73E2
<NRf+>
Expanded decimal format that includes <NRf>, MIN and MAX. Examples: 273
273. 2.73E2 MAX. MIN and MAX are the minimum and maximum limit values
that are implicit in the range specification for the parameter.
Table 6-2. Suffixes and Multipliers
Class Suffix Unit Unit with Multiplier
Current A Ampere MA (milliampere)
Amplitude V Volt MV (millivolt)
Time S second MS (millisecond)
Common Multipliers
lE3 K kilo
1E-3 M milli
1E-6 U micro
Boolean Data
Either form {1|0} or {ON|OFF} may be sent with commands. Queries always return 1 or 0.
OUTPut OFF
CURRent:PROTection 1
Character Data
For query statements, character strings may be returned in either of the forms shown in Table 6-3, depending on the length
of the returned string.
Table 6-3. Character Data Formats
<CRD>
Character Response Data. Permits the return of character strings.
<AARD>
Arbitrary ASCII Response Data. Permits the return of undelimited 7-bit ASCII. This data type has
an implied message terminator.
Note:
The IEEE 488.2 format for a string parameter requires that the string be enclosed within either single
(’ ’) or double (" ") quotes. Be certain that your program statements comply with this requirement.