2.6.1 X-STREAM X2GK, X-STREAM X2GP
2.6 Installation - X-STREAM X2GK, X-STREAM X2GP
Fig. 2-4: X-STREAM X2GK - rear panel
Plugs and sockets required for the electrical
connections are on the rear panel of the units
( Fig. 2-4 & Fig. 2-5 ).
X-STREAM X2GP analyzers provide an in-
ternal wide range power supply for worldwide
use.
X-STREAM X2GK analyzers are powered by
an internal wide range power supply for world-
wide use, or by an external DC 24 V power
supply unit, optionally supplied with the unit.
Gas inlets
Gas outlets
AC Power inlet
with fuses
Analog outputs / relay
outputs
RJ45 ethernet
connectors
Serial interface
( RS 232 /
Modbus 485)
Digital in- /
outputs
Service interface
If an external PSU is not included in delivery,
another unit can be used instead, provided it
conforms to the specications given by the
X-STREAM X2 instruction manual.
X-STREAM X2GK / X2GP analyzers should
be operated in a horizontal position.
Six screws at the front panel enable to install
X2GK models into a rack. The external PSU
is optionally available for rack mounting, too.
X2GP can be installed into a rack by adding
two optional brackets to the left and right hand
side of the instrument.
Alternative
DC Power inlet
and separate
fuse
HAZARD BY RACK INSTALLATION
The front panel and its screws
are not designed to carry the
weight of the instrument!
Support the instrument, when
installed into a rack!
Disregarding may cause per-
sonal injury and damaged
equipment.