Startup
3.7 Calibrating the Analyzer
3.7.1 Preparing Calibrations
Before starting calibrations it is required to
tell the instrument the calibration gas concen-
trations.
Starting from the measurement screen press
the DOWN key to open the MAIN MENU,
enter the SETUP-CALIBRATION.. menu and
directly enter the CALIBRATION GASES..
menu.
Calibration gases
ZeroGas: 0.0 ppm
SpanGas: 500.0 ppm
Enter the concentration value for the zero gas
to be used during zero calibration.
Enter the concentration value for the span
gas to be used during span calibration.
Note!
The units for the calibration gases are taken
from the related entry in the display setup
menu.
When done, press the key to return to the
CALIBRATION menu.
Setup..
Calibration..
Calibration gases..
Multi-channel unit:
Select the channel to be calibrated in the
SELECT COMPONENT menu.
Multi-channel unit:
Press the key to enter the SELECT COM-
PONENT menu to change the settings for a
different channel.
Consider the calibration notes
in section "Measurement spe-
cications"!