Emerson Process Management X-STREAM X2 Series Welding Consumables User Manual


 
Emerson Process Management GmbH & Co. OHG2-18
X-STREAM X2
Short Form Manual
HASX2E-SFM-HS
02/2012
All signal cables are connected to screw-type
terminals located inside the housing. Access
to the internal components is gained by re-
leasing the two fasteners at the top of the unit
and opening the front panel downwards.
All cables must be fed through cable glands
and secured with a gland nut.
Preparation of signal cables
Gas connections
2.6 Installation - X-STREAM X2XF Field Housings
Installing cable glands with shielded cables
3. Feed cable through
dome nut and clam-
ping insert
4. Fold braided shield
over clamping insert
5. Make sure that brai-
ded shield overlaps
the O-ring by 2 mm
(
3
32“)
1. Strip the cable
2. Expose braided
shield
6. Push clamping insert
into body and tighten
dome nut
7.
Assemble into hou-
sing and you´re done!
Gas connections
max number 8
max for purging (incl. / separate) 2 incl.
material PVDF; stainless steel (opt.)
sizes 6/4 mm;
1
4"
Signal in- and outputs
Properly installed, the glands act as a strain
relief and guarantee EMC (electromagnetic
compatibility):