page 116 008-0608-00
ity.
•“
KNOWN POINT 2/3
”:
point 2 of 3, usually 50% of the transducer’s full-scale capac-
ity.
•“
KNOWN POINT 3/3
”:
point 3 of 3, usually 100% of the transducer’s full-scale
capacity.
When using the 5-Point Known Load Calibration type, the following menu items
are available:
•“
KNOWN POINT 1/5
”: point 1 of 5, usually 0% of the transducer’s full-scale capac-
ity.
•“
KNOWN POINT 2/5
”:
point 2 of 5, usually 25% of the transducer’s full-scale capac-
ity.
•“
KNOWN POINT 3/5
”:
point 3 of 5, usually 50% of the transducer’s full-scale capac-
ity.
•“
KNOWN POINT 4/5
”:
point 4 of 5, usually 75% of the transducer’s full-scale capac-
ity.
•“
KNOWN POINT 5/5
”:
point 5 of 5, usually 100% of the transducer’s full-scale
capacity.
These menu items are only available with the Known-Load Calibration types.
NOTICE
To insure both correct operation of the transducer and application of the load,
the instrument expects the voltage or current applied at each known-load point
to be increasing. For example, the load applied at Known-Load Point 2/2 must
cause the transducer to produce a more positive voltage or current than at
Known-Load Point 1/2.