Honeywell SC1000 Stud Sensor User Manual


 
SC Series Instruction Manual page 153
Error Messages17
Error 29,
S.EE VERIFY FAIL
A verification failure occurred during a write operation to the channel’s EEPROM.
Restart the instrument.
Error 30,
S.EE ERASE NAK
The channel’s EEPROM could not be detected during an erase operation. Check
that the channel’s address jumpers are set properly.
Error 31,
BLANK IDENT
The channel does not appear to have been installed properly. This occurs if you
swap Input or Output boards without using
SYSTEM MENU
->
INSTALL CHANNEL
to format
the channel’s EEPROM.
Error 32,
BOGUS EE ADDR
An invalid address has been passed to the programmer. Restart the instrument. If
this problem persists, contact our Service department with the information found
in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 33,
SIG.MOD WIRES
The instrument has detected that the (+)Signature and (-)Signature pins (labeled
as “+MEM” and “-MEM) are likely reversed.
Error 34,
SIG.MOD MISMATCH
The excitation type byte found inside a Signature Module indicates that the chan-
nel is not designed to operate with that type of transducer.
Error 35,
BOGUS MENU TYPE
A menu contains an invalid menu type. Contact Sensotec with the information
found in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 36,
BAD EE TABLE
The instrument doesn’t understand the EEPROM type setting for the channel.
Contact Sensotec with the information found in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 37,
BAD PERSONALITY
The channel doesn’t understand the given card type. Contact Sensotec with the
information found in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 38,
BAD FORMAT
The channel’s EEPROM memory contains a display setting which is not under-
stood by the instrument. Check that every item in the channel’s
DISPLAY SETUP
menu has a valid setting.
Error 39,
WRONG HARDWARE
The channel type expected by the instrument for the channel number is not the
channel type that is physically installed. If you have changed the address jumpers
on one or more channels, replace them in their original configuration.
Error 41,
SHUNT CAL ERROR
The application of the shunt resistor did not affect the transducer’s output. Check
the wiring and the transducer output with a voltmeter to verify proper operation.
If this is a Strain Gage Channel, make certain that the shunt calibration resistor
installed on the circuit board is the correct value per the transducer’s Calibration
Record.
If this is a High-Level Input Channel, make certain that the Shunt Cal 1 and Shunt
Cal 2 pins are connected to the correct pins of the transducer/amplifier.
Error 43,
BAD IDENT CRC
A checksum failure has occurred when reading a channel’s EEPROM. Restart
the instrument.
Error 44,
BAD FUNCTION
The instrument has requested that the channel perform a function that it is not
capable of. Restart the instrument.
Error 45,
A/D NAK
Hardware error. The channel’s Analog-to-Digital converter is not communicating
with the channel’s microprocessor. Restart the instrument.
Error 46,
SENSE WIRES
The Strain Gage Channel’s (+)Sense and (-)Sense pins (labeled as “+SEN” and “-