3M #3968 Stud Sensor User Manual


 
3M™ Rapid Detection Reader Managing Settings
21
Settings: View Tabs
Tab Screen
Timer continued
NOTE: Only new results that have not been previously
transferred are transferred during an upload. Data is not deleted
from CM upon transfer. When the CM database is full, the oldest
result is over-written.
NOTE: It is important that the upload interval be set to ensure
data capture and transfer prior to records being overwritten.
Auto IQC Interval
Edit on CM and through RCU.
Purpose: The interval between automatic IQC self-diagnostics in
either local or networked environment. Disable by entering 0.
Range: 0 1440 minutes.
Default: 0 (Disabled).
Printer
Edit through RCU only.
Auto Print
Purpose: Defines if automatic printing is enabled. Settings are
independent for assay, LQC and IQC results.
On: Result will automatically print upon completion.
Off: Result will not print automatically.
Default: Off.
Print Headers
Purpose: Define customized print header information that will be
visible on every printout.
Figure 24. Settings: View Printer
Tab
Login
Edit through RCU only.
Device Name
Purpose: Define name assigned to the device.
Default: Blank.
Lockout
Purpose: Disable Reader.
Device Locked: Assays cannot be run.
Device Unlocked: Assays can be run (Default).
Figure 25. Settings: View Login
Tab