3M #3968 Stud Sensor User Manual


 
3M™ Rapid Detection Reader Managing Settings
25
Settings: Edit Tabs
Tab Screen
Date/Time
Edit on CM. Can only edit through RCU if NTP is enabled.
Purpose: Set the current date and time for the 3M Rapid
Detection Reader.
If in Auto transfer Mode, the 3M Rapid Detection Reader date and
time may be synchronized with the NTP server. The date format
in the Reader is variable.
To Change:
1. Use drop down menu to scroll through times zones and select
correct time zone offset for your location.
2. Enter Current time.
NOTE: When adjusting for changes in daylight savings time
ensure the time zone offset is adjusted not only the current
time.
NOTE: After changing the date or time, the CM power must be
cycled to reset auto-IQC and data transfer timers.
NOTE: The date format in the RCU is the same as for other
applications running on that PC.
NOTE: Periodically check that the clock is correct and has not
been inadvertently changed.
Figure 30. Settings: Edit
Date/Time Tab
Language
Edit on CM only.
Purpose: Set the display language for the 3M Rapid Detection
Reader.
Figure 31. Settings: Edit
Language Tab
Misc.
Manual / Auto Transfer Mode
Edit on CM. Cannot edit through RCU.
Manual Transfer Mode: 3M Rapid Detection Reader is set up for
stand-alone operation.
Auto Transfer Mode: 3M Rapid Detection Reader is set up for
network operation.
Default: Manual Transfer Mode.
Figure 32. Settings: Edit Misc. Tab