3M™ Rapid Detection Reader Maintenance
36
Maintenance
WARNING!
To reduce the risk associated with hazardous voltage:
• Unplug the 3M Rapid Detection Reader before cleaning.
• Do not disassemble the 3M Rapid Detection Reader or any of its related
components and accessories. The instrument contains no operator-serviceable
components.
• Do not immerse any of the 3M Rapid Detection Reader components in any liquid.
• Do not use the 3M Rapid Detection Reader if there is apparent damage to a power
cord or power supply.
To reduce the risk associated with potentially infectious patient samples:
• Do not spill specimen or sample fluids on any of the 3M Rapid Detection Reader
components or on the outside of the Test Cartridge.
• If a spill occurs, disinfect external surfaces only, using a soft cloth containing a
solution of 0.5% bleach, 70% isopropyl alcohol or 70% ethanol.
The 3M Rapid Detection Reader requires no regular maintenance or calibration during its
intended life span, other than exterior cleaning as needed.
Any liquids on the external surface of a Test Cartridge must be cleaned before inserting into the
ports of the TM.
Exterior Cleaning
CAUTION.
To reduce the risk of instrument or accessory damage:
• Do not immerse the 3M Rapid Detection Reader or related devices in any liquid.
• Do not spill liquids on the 3M Rapid Detection Reader and related accessory
devices.
• Do not expose the 3M Rapid Detection Reader to temperatures above 60°C
(140°F). Do not disinfect or sterilize the 3M Rapid Detection Reader using devices
such as autoclaves, ultrasonic cleaners, or pasteurizers. Use of such methods to
disinfect/sterilize the 3M Rapid Detection Reader will cause permanent device
damage and void the warranty.
The external surfaces of the TM and CM can only be disinfected using a solution of 0.5% bleach,
70% isopropyl alcohol or 70% ethanol applied to a soft cloth.
CAUTION.
Possible device damage:
• DO NOT USE STRONG BLEACH (>0.5% solution) OR ACETONE! Oxidants and
solvents could damage the 3M Rapid Detection Reader casing and touch screen.
• Clean external surfaces only.