ENTRY OF SYSTEM PARAMETERS
ETC00265(4) Series 100 e 02/2004
OFFSET (BEGIN OF RANGE)
8.12 Offset (Begin of range)
The operator is here offered the opportunity to introduce a scale offset for the analog signal output
(begin of range).
Example:
For an analyzer concentration range of 0 - 25 % it is desired to measure only concentrations in
the range 10 - 25 %. If the operator enters here the value 10 %, the analog signal outputs of
0 V / 0 mA or 2 (0.2) V / 4 mA will then correspond to a gas concentration of 10 %.
The displayed values are not affected.
Effect the entry using
and
Entry possibility for channel 2
Use the keys
and for the entry.
Note:
The specifications of the analyzer written in the data sheet are only for OFS. = 0 and
END = full - scale range set in our factory !
It is part of customer to enter logical values for OFS. and END !