Emerson Process Management BINOS 100 M BINOS 100 2M Welding Consumables User Manual


 
TECHNICAL DATA
27 - 8
ETC00781(4) Series 100 e 02/2004
GENERAL SPECIFICATIONS
27.5 Voltage Supply
BINOS
®
100 (M) / OXYNOS
®
100 / HYDROS
®
100 / BINOS
®
100 2M (external power supply)
Input 3-pole XLR- Flange (male), lockable
Voltage Supply 24 V DC (+/- 5 %)
[DC voltage supplied by optional power supplies SL5, SL10 (both for rack mounting only) or UPS 01 T
or equivalent power supply]
Power Consumption (analyzer itself) BINOS
®
100 (M) < 20 W
OXYNOS
®
100 < 40 W
HYDROS
®
100 < 35 W
BINOS
®
100 2M < 120 W
BINOS
®
100 2M (internal power supply)
Input plug
integrated power supply UPS (input voltage see chapter 27.5.2)
Power Consumption max. 240 VA
Output to external consumers 3-poliger XLR- Flange (female)
Available power supply to max. 2.0 A (depends on internal configuration)
external consumers
BINOS
®
100 F
Input terminal strips
Internal power supply SL5 (input voltage see chapter 27.5.2)
Power consumption max. 350 VA
Fuses (internal) T 3.15 A / 250 V (2 pieces)
27.5.1 Electrical Safety
Over-voltage category II
Pollution degree 2
Safety Class 2 for BINOS
®
100 (M)/ OXYNOS
®
100 /
HYDROS
®
100/BINOS
®
100 2M (ext. PS)
instruments.
1 for BINOS
®
100 2M (int. PS)/ BINOS
®
100 F
instruments
all I/O´s SELV voltage
optically isolated to electrical supply