CALIBRATION
9 - 10
ETC00781(4) Series 100 e 02/2004
9.3 Remote-Controlled Calibration Mode (Option)
A remote-controlled calibration requires internal solenoid valves or external solenoid valves
controled via digital outputs. Additional the correct test gas setpoint has to be set first (manually:
chapter 9.1.2 / serial interface option chapter 12.5). The automatic function of the analyzer must
also be activated correctly (cf. chapter 8.5).
A calibration can be started via
P serial interface option (chapter 12.5)
P digital inputs option (chapter 13.1)
P FOUNDATION
Fieldbus (chapter 13.2)
A calibration will be performed as described in either chapter 9.1 (single channel zeroing or
spanning) or in chapter 9.2.2 (combined zeroing and spanning), depending on start command.