Error Code Possible Reasons Check / Correct
16 - 6
LIST OF FAILURES
ETC00781(4) Series 100 e 02/2004
9. Barometric pressure
effects.
10. Temperature below the
dew point in the gas paths.
11. Faulty A/D converter.
1. Incorrect response time
( t
90
time) settings.
2. Pumping rate inadequate.
3. Contamination of the gas
paths.
9. Enter the correct value for
barometric pressure (see chapter 8.1).
Pressure sensor faulty (E.37).
10. Check the temperature of the gas
paths and eliminate any reason of
condensation,
Maintain all temperatures at values
at least 10 °C above the dew point of
sample gas.
11. Exchange BKS.
1. Check the value for t
90
time
(see chapter 8.11).
2. The feeder line between the
sampling point and the analyzer is
too long.
Use a larger, external pump;
consider adding a bypass line to the
process stream for sampling
purposes (see chapter 5.1).
3. Check gas paths and gas
conditionning to contamination.
Clean gas paths and exchange the
filter elements.
Fluctuating or
erroneous display
Response - time too long
(t
90
- time)