8 - 14
ENTRY OF SYSTEM PARAMETERS
ETC00265(4) Series 100 e 02/2004
8.17 Pump
*)
This display shows the status of gas sampling pump.
If the function is enabled in “Pump Control” (“P.Ctl.”, see chapter 8.18), the pump can be switching
On or Off via this input.
Display / Entry of 0: gas sampling pump is Off
Display / Entry of 1: gas sampling pump is On
Entry is performed using
followed by
8.18 Pump Control
*)
In this item you can determine the input for pump control.
Entry of 0: pump control via front panel key
Entry of 1: pump control via system parameter “PUMP” (chapter 8.17)
Entry of 2: pump control via optional digital input (Item 13.1)
Entry is performed using
followed by
Push the key until
the displays show
The analyzer now is back in the analysis mode.
GAS SAMPLING PUMP
*)
BINOS
®
100 2M/F with software version 5.0 or higher only