Emerson Process Management BINOS 100 M BINOS 100 2M Welding Consumables User Manual


 
TECHNICAL DATA
27 - 1
ETC00781(4) Series 100 e 02/2004
27. Technical Data
Certifications EN 61326, EN 61010-1
(for measurment of non flammable gases or non
explosive gases resp. (< 50 % LEL).
NAMUR, CSA-C/US,
Higher concentrations require supplementary C-Tick, PAC, GOSSTANDART
protective measures !)
Suitability tests TÜV Nord mbH, report no.: 95CU054/B,
TÜV Bavaria, report no.: 1563793,
Carbon monoxide (CO): TI Air, 13
th
BImSchV
Oxygen (O
2
): TI Air, 13
th
BImSchV, 17
th
BImSchV
DMT: IBS/PFG-No. 41300392NIII/41300292NIII
CO
2
/ CH
4
: 0-80 Vol.-%
CO : 0-200 ppm / 0-10 Vol.-%
O
2
: 0-10 Vol.-%
For information about the configuration, measuring components and ranges of the actual
analyzer refer to the nameplate label, which is located on the left or right side of the analyzer
cover (all table top and rack mount versions) or inside the front door (BINOS 100 F).
How to read a name plate label:
27.1 Options
Pressure sensor Measuring range 800 - 1,100 hPa
Fine dust filter (BINOS
®
100 2M/F) Filter material PTFE, Pore size approx. 2 mm
Sample gas pump (BINOS
®
100 2M/F) Pumping rate max. 2.5 l/min.
Suction pressure min. 900 hPa
Lifetime max. 5,000 hours
Channel 1:
Component and
full scale range
Channel 2:
Component and
full scale range
Analyzer type and
measuring prinziples
Serial number
Analog output
settings
DMT approval number