Emerson Process Management BINOS 100 M BINOS 100 2M Welding Consumables User Manual


 
SAFETY SUMMARY
S - 5
ETC00781(4) Series 100 e 02/2004
X BINOS
®
100 2M (internal PS) and BINOS
®
100 F are Safety Class 1 instruments
The analyzer is provided with a protective earth terminal.
To prevent shock hazard, the instrument chassis and cabinet must be connected
to an electrical ground. The instrument must be connected to the AC power
supply mains through a three-conductor power cable, with the third wire firmly
connected to an electrical ground (safety ground) at the power outlet.
If the instrument is to be energized via an external power supply, that goes for the
power supply too.
Any interruption of the protective (grounding) conductor or disconnection of the
protective earth terminal will cause a potential shock hazard that could result in
personal injury. Deliberate disconnection is inadmissible / prohibited !
The analyzer BINOS
®
100 F (field housing) has no switch with disconnect
function. The customer has to provide a switch or circuit breaker into his
installation. This switch has to be installed near by analyzer, must be easily
attainable for operator and has to be characterized as disconnector for analyzer.
Cables to external data processing have to be double-insulated against mains
voltage for analyzer BINOS
®
100 F !
Use cables suitable for intrinsic safe applications only ! Install internal data lines
that they have a distance to mains voltage lines of at least 5 mm.
This distance has to be valid permanently (e.g. via cable holder) !
24 VDC supply to external components/analyzers with the internal power
supply of BINOS
®
100 2M requires a fuse to be connected in series to the
consumer which limits the current consumption to max. 2 A !
Verify correct polarity for 24 V DC supply of external components !
SUPPLY VOLTAGE