Emerson Process Management BINOS 100 M BINOS 100 2M Welding Consumables User Manual


 
6 - 4
SWITCHING ON
ETC00781(4) Series 100 e 02/2004
6.3.1 BINOS
®
100 2M
The analyzer BINOS
®
100 2M is optional lly equipped with an internal “autoranging” power
supply and is specified for an operating voltage of 230 V AC or 120 V AC resp., 47-63 Hz.
For supply of external components / analyzers there is built in a 24 VDC outlet (a 3-pole XLR flange,
female, max. 2 A).
P Connect internal power supply and external components (Fig. 6-2, 24 V DC out).
24 VDC supply to external components/analyzers with the internal power
supply of BINOS
®
100 2M requires a fuse to be connected in series to the
consumer which limits the current consumption to max. 2 A !
Verify correct polarity for 24 V DC supply to external components before
operation (Section 29.) !
P Connect mains line and internal power supply (Fig. 6-2, V AC in).
Verify beforehand that the line voltage stated on the power supply meets that
of your power supply line !
The socket outlet shall be installed near the equipment.
230/120 V AC SUPPLY