12.5 Specifications
(1)
Total excitation current of all channels must not exceed 120mA for Models SC1000,
SC2000 or SC2001 and 310 mA for Model SC3004.
TRANSDUCER INPUT
Transducer types amplified with voltage or current output,
(field selectable)
Excitation Voltage
(1)
±15 VDC @ 40 mA max.
+28 VDC @ 75 mA max.
+12 VDC @ 40 mA max.
(field selectable)
Excitation Overcurrent Protection yes
Transducer full-scale output ±1, ±5, ±10 VDC
4-20 mA (field selectable)
Calibration Type shunt, or
2-, 3- or 5- point known load
A/D Converter 24-bit Sigma-Delta
Low-pass filter digital, 24-tap FIR
Resolution and Frequency Response see “
FREQ. RESPONSE
Menu Item” on
page 109
INSTRUMENT-ONLY ACCURACY
(Frequency Response setting<=16Hz; Linearity, repeat-
ability & hysteresis)
Known -Load Calibration ±0.01% F.S.
AUXILLIARY INPUTS
Quantity 2
Type momentary contact closure
Response Time < 5ms
Field-Selectable Functions tare on, tare off,
peak/valley clear, peak/valley hold,
track hold
ANALOG OUTPUT
Output voltage range 5, ±5, 10 or ±10 VDC (field selectable)
Output current range (optional current
output channels)
4-20 mA
Source any channel’s track, peak or valley value
Isolation 500V
Resolution 13 bits
Frequency Response same as input when driven by the same
channel’s tracking data